From: "Meditsiiniseadmed (Medical Devices)" <[email protected]>
Sent: Tue, 02 Apr 2024 10:20:00 +0000
To: "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "[email protected]" <[email protected]>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "[email protected]" <[email protected]>; "'[email protected]'" <'[email protected]'>; "[email protected]" <[email protected]>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "[email protected]" <[email protected]>; "[email protected]" <[email protected]>; "'[email protected]'" <'[email protected]'>; Medisinsk utstyr <[email protected]>; "[email protected]" <[email protected]>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; JAZMP MP_SFI <[email protected]>; JAZMP Meddev <[email protected]>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "'[email protected]'" <'[email protected]'>; "Medical Devices at Medicines Authority" <[email protected]>; Katja MohorÄiÄ <[email protected]>; "[email protected]" <[email protected]>; "[email protected]" <[email protected]>
Subject: EE Derogation survey results
Dear Colleagues,
We would like to thank the 18 countries that participated in the derogations survey organised by the Estonian competent authority.
The responses to the survey are crucial for us to enhance national derogations procedures and develop forms.
We are pleased to share the results of the survey.
Please find the attached file with the results.
Kind regards,
Maarika Ojala
Chief Specialist
Department of Medical Devices
+372 5400 4028
Republic of Estonia
Health Board
+372 794 3500
www.terviseamet.ee
Paldiski mnt 81, 10614 Tallinn
Estonia
This e-mail is confidential and meant for use by the person named in the letterhead. Any use in any way or copying of it by a person not marked as the addressee thereof is prohibited. If you have got this e-mail by mistake, please notify of it the sender without delay and delete the received e-mail together with all its attachments.
Saatja: Maarika Ojala <[email protected]>
Saatmisaeg: reede, 2. veebruar 2024 08:30
Adressaat: '[email protected]'; '[email protected]'; '[email protected]'; '[email protected]'; '[email protected]'; '[email protected]'; '[email protected]'; [email protected]; '[email protected]'; '[email protected]'; '[email protected]'; '[email protected]'; '[email protected]'; '[email protected]'; '[email protected]'; [email protected]; '[email protected]'; [email protected]; '[email protected]'; '[email protected]'; [email protected]; [email protected]; '[email protected]'; Medisinsk utstyr <[email protected]>; [email protected]; '[email protected]'; '[email protected]'; JAZMP MP_SFI <[email protected]>; JAZMP Meddev <[email protected]>; '[email protected]'; '[email protected]'; '[email protected]'; '[email protected]'; Medical Devices at Medicines Authority <[email protected]>; Katja Mohorčič <[email protected]>; [email protected]; [email protected]
Teema: REMINDER Invitation to participate in a survey on derogations (national and compassionate/single patient use)
Dear colleagues,
We would like to keep you updated on the progress of the study on derogations and reiterate our call for your participation.
In particular, we would like to thank the 7 countries that have already responded.
The survey is open until 29.02.2024 and we hope that more colleagues will find the time to fill in the questionnaire:
https://ec.europa.eu/eusurvey/runner/EEDEROSurvey2023
We would be very grateful if you could pass this link on to those colleagues who deal with derogations.
If you have any questions, please do not hesitate to ask and send an email to [email protected]
Best regards,
Maarika Ojala
Chief Specialist
Department of Medical Devices
+372 5400 4028
Republic of Estonia
Health Board
+372 794 3500
Paldiski mnt 81, 10614 Tallinn
Estonia
This e-mail is confidential and meant for use by the person named in the letterhead. Any use in any way or copying of it by a person not marked as the addressee thereof is prohibited. If you have got this e-mail by mistake, please notify of it the sender without delay and delete the received e-mail together with all its attachments.
EEDEROSurvey2023
Derogation from the conformity assessment procedure
29.02.2024
Department of Medical Devices
Estonian Health Board
Maarika Ojala
Chief Specialist
Background
Given the complexity of dealing with derogations from the conformity
assessment procedure established with Article 59 of the MDR (EU) 2017/745
and Article 54 of the IVDR (EU) 2017/746, it was planned to carry out a study
in order to establish procedures and develop application templates
in Estonia.
AIM
How have the procedures for these two different derogations been
established in other member states?
What are the countries' experiences in processing
derogations?
Content
National derogation
Compassionate use/single patient derogation
Further information experience, comments
General information country, institution, general e-mail address of the entity(-ies) processing the derogation
Future plan regarding derogations
Results
19 countries participated
The survey lasted between
16.01-29.02.2024
National derogation MDR 2017/745 Article 59 IVDR 2017/746 Article 54
Is the procedure for making derogation laid down in national law?
Do you have standard application/request templates?
Who can apply for a derogation?
Someone else who can apply • Ministry of Defence may derogate from the Act on MDs but only as a distributor or service person
• Everybody can apply for an exceptional derogation but only the manufacturer or owner of the device can receive the approval
• Who wants is not determined
• Legally everybody, practically only with the cooperation of the manufacturer
• Dual application from manufacturer/authorized rep and healthcare professional needs to be submitted, the healthcare professional
application should be completed by an appropriately senior clinician, who is a representative of the relevant health service being provided,
e.g. at institutional or national level
• If other economic operators other than the manufacturer apply, they must submit the authorization certificate obtained from the
manufacturers that they represent in order to apply for derogation
• All the other if authorized by the manufacturer
• The derogation is always granted to the manufacturer or its authorized representative, even if it is the distributor who initiates the request
• An application for an exemption authorization can be submitted by anyone who is responsible for placing the device on the market or
putting it into service
• It is not determined by law who is allowed to apply for a derogation
• If it is not the manufacturer, we like to see a proxy, but we see the manufacturer as the one who can apply
• Any other person duly mandated by the manufacturer or its authorized representative (a contract signed by both parties must prove this
mandate)
In summary: who can apply
• There is no provision on who can apply
• Derogation granted to the manufacturer or EU-REP
• The manufacturer must be linked to the derogation
When you ask about the reasons for the derogation, do you ask...
Other reasons for derogation • Suitable alternatives available
• Medical expertise, a statement of the NB, conformity to safety and performance requirements, approval or certification in other industrial states worldwide, experiences in human/animal studies, current completed or planed clinical trials
• An estimate of the affected patient population, similar derogation, an overview of the CA-s incl decision and reasoning, expected date that the CE certificate will be issued and timeframe confirmed by the NB, stock situation in combination with usage figures resulting in the number of days of stock at the most critical care providers, an overview of the customers with whom these risks will occur, timeline in which these risks are expected, an explanation of whether the MD can replace in short term by other MD from other MF incl explanations, a confirmation that the products covered by this application for derogation have not been adapted/changed after the CE certificate has expired and will be produced and delivered under exactly the same conditions in the event of a possible derogation, if CE certificate has already expired, proof that customers/healthcare institutions have been informed about the invalidity of the CE certificate
• The medical reason for the need of the device, amount; alternatives, previous CE status, etc
• MDR transition/notified body status, but this is not a prerequisite for an A59 derogation
• If device marketing/use/available outside the EU/EEA, number of devices, specification on when the device is expected to be CE marked, if the device has been used before in clinical trials or performance studies, how the device differs from other similar devices already on the market, explanation on the documentation that provides support for the device to be used safely and correct for both the patient and other users, details on vigilance or recall issues to date, indication on what measures have been taken to find an alternative device that meets the requirements of the regulations, explanation why any available CE-marked device(s) or are not a suitable alternative for the patient(s), describe possible consequences if the patient(s) cannot access the device(s)
Other reasons for derogation • Statement that the device has no alternative on the national market, device does not pose an unacceptable risk to public health and patient
safety, evidence that the device fulfills the general safety and performance requirements, taken into account the intended use, (if available) the market authorisation status of the device in other countries, written opinions of health institutions and organisations or their competent authorities on the neccesity of the device, including the reasons why the device cannot be supplied
• The number of medical devices to be supplied under the exemption and the list of end-users, documentation demonstrating the need for placing the device on the market, need for the medical device concerned, evidence that there is no equivalent alternative on the market (or equivalent device or other treatment), an explanation of how the device concerned differs from others on the market, market share, availability of the device, information on the regulatory status of the medical device, certificates, notifications, declarations valid outside the EU (certified copies of documents), information on regulatory status in other EU other Member States have granted/refused derogation or still in the process – not yet completed, conformity history of the device – whether the device had CE marking, certificates, declarations in the EU before the new Regulation came into force, whether the device is intended for certification for the EU market or is already in the process of certification (evidence that the manufacturer and the notified body have signed the contact, and that the assessment procedure is ongoing), whether critical parts of the device have already been assessed and demonstrate an adequate level of safety for the user or recipient (patient), confirmation general safety and performance requirements, data on clinical studies/testing, an indication when certification under the MDR by a notified body is supposed to be concluded
• Applicant provides all the safety guarantees
• Justification as to why the conformity assessment procedure has not been completed, details incl proof of the planned completion of the conformity assessment procedure, evaluation of potential alternative devices and/or therapeutic procedures, justification as to why the placing on the market/putting into service of the devices are in the interests of public health or patient safety or health, number of devices placed on market in the last 3 years, estimated number of MD under the derogation, list of customer, proof general safety and performance requirements
In summary: other reasons for derogation
• Alternative devices or treatments
• General safety and performance requirements
• Statement of the Notified Body
• Differs explanation from others on the market
• Data on clinical studies/trials/performance studies
• List of customers
• Affected patient population
• Stock, market situation
What supporting documents do you ask for...
Other supporting documents • Documentation based on provisions laid down in Annex II, III and IV of Regulation (EU) 2017/745, up-to-date and
complete to the maximum extent incl. a justification as to why a particular part of the documentation was not submitted
• Risk and vigilance assessment
• One or more user(s) or healthcare provider(s) to explain the effect on patient care or patient safety if the derogation is not granted and to declare that no equivalent
• Risk analysis, countries where placed on market, statement from healthcare professional
• Full technical documentation is not requested in all circumstances
• Statement that the device has no alternative on the national market, statement that the device does not pose an unacceptable risk to public health and patient safety, evidence that the device fulfills the general safety and performance requirements, taken into account the intended use, (if available) the market authorization status of the device in other countries, (if available) written opinions of health institutions and organisations or their competent authorities on the neccesity of the device, including the reasons why the device cannot be supplied
• Additional requests for information on case by case
• Case by case, request the latest EC declaration of conformity issued by the manufacturer and we request a declaration of conformity to the general requirements, part of the technical documentation may be requested, sales history-especially history of use, in certain cases, it may be asked to identify current certification problems
Other supporting documents • Regarding the number of devices depends if derogation is granted on number of units or with a timeframe
• It depends on the situation/product what we ask for
• Any decisions already given by other EU member states/marketing authorisations in other jurisdictions, if the
certificate is still valid, plan to put devices on the market before the certificate expires in order to build up stocks?
explanation and details, if applicable, of alternatives for each device, for a device previously placed on the market,
the list of customers (name, address) and the list of the number of devices ordered for each customer (name and
reference) over the last 3 years, a proof that there is a quality management system in place, provide the date of the
last surveillance audit report, confirmation letter by the notified body that application for MDR certification has been
accepted and the contract signed with the manufacturer, commitment by the notified body to inform the CA about
major safety-related shortcomings identified during conformity assessment, a detailed roadmap for obtaining the new
certificate, reinstating the suspended certificate (if applicable), provide a report containing relevant data gathered
through its post-market surveillance (PMS) system, in particular data concerning incidents, serious incidents and/or
field safety corrective actions, should based on these elements the benefit-risk ratio is still positive, provide
information about potential safety-related shortcomings identified by notified body during last surveillance audit and
confirmation regarding satisfactory resolution
In summary – conclusions on the reasons for derogations and supporting documents
• It appears like duplicate questions for respondents
• Question of supporting documents was presented in general terms, but the respondents
have gone into detail in their answers, presented the content that is asked in the following
separately questions
• It appears that the entire tehnical documentation is not being requested
• The procedure is carried out on a case-by-case basis
Do you ask about similar applications in other EU countries?
Does the healthcare institution/professional user also justify the derogation?
Do you also ask the applicant to confirm or take responsibility that it is a non-compliant medical device?
Approximately how many national derogation requests do you receive/approve/reject per year?
Country Receive Approve Reject
1. ca 14 Ca 7 Ca 7
2. 5-10 4-9 (ca 90%) 1-2 (ca 10%)
3. - - -
4. 0 0 0
5. 20 16 3-4
6. 12-13 10 2-3
7. 10 8 2
8. 10-15 10-15 0
9. 4 0 4
10. 15 10 0
Approximately how many national derogation requests do you receive/approve/reject per year?
Country Receive Approve Reject
11. 5 2 3
12. 3-4 3-4 0-1
13. 50 29 8
14. 1 1 1
15. 10 9 1
16. 3 3 0
17. 20-30 23 2
18. 15 13 2
19. Annual report 2022
26 13 ?
If rejected, what is the most common reason?
Other reason
• No support of healthcare professional
• There is no safety information on the device (device not yet marketed)
• Mainly for failure to provide the necessary documents
• That on a overall assessment the interest of public health or patient safety or health cannot
be justified
• Device intended for clinical study
• Lack of positive benefit/risk data
Have you had any cases where the applicant has withdrawn the application or failed to complete the process?
Reason for withdrawal or failure to complete process • Perhaps because the device was unsafe
• Devices have availed of amended transitional provisions to MDR
• Transitional provisions, change of market situation
• Withdrawn as a result of an EU exemption or the applicant was not entitled to apply for an exemption
• The applicant is not able to fullfil the requirements
• Unknown
• Not sufficient supporting documents
• Adoption of Reg 2023/607, prolonged validity of MDD certificate
• When the request comes from the healthcare professional but the manufacturer does not wish to follow this procedure, or
when the applicant agrees to follow our proposal to set up a clinical investigation
• Failure to submit sufficient evidence
• Costs
• The reasons are different for every case but it happens when sending them a first round of questions, assuming that they
realize that they don't comply with the derogation requirements
Fees Country Fee amount
1. 250-10.300 EUR (acc. to nat. fee ordinance)
2. Not yet
3. It is written in the national law that we can ask for a processing fee, however we are currently not asking for a derogation processing fee
4. 4000e, which may be waived at CA's discretion
5. 1.084,67 Euro
6. The fee varies between 1930€ - 4070 €
7. There's no fee for derogation procedure, just fee for General administrative procedure (22,66 EUR)
8. the hourly price is 200 francs
9. 1000 euro
10. 193 Euro per hour
11. Not yet
12. 62000 SEK
Compassionate use/single patient derogation MDR 2017/745 Article 59 IVDR 2017/746 Article 54
Is the procedure for making derogation laid down in national law?
Do you have standard application/request templates?
Who can apply for a derogation?
Someone else who can apply
• Who wants is not determined
• Dual application from manufacturer/authorized rep and appropriate specialist needs to be
submitted
• All the other if authorized by the manufacturer
• The healthcare professional justifies the public health interest
• It is not determined by law who is allowed to apply for a derogation
• Any person duly mandated by the manufacturer or authorized representative (a copy of the
mandate has to be provided)
When you ask about the reasons for the derogation, do you ask...
Other reasons for derogation • Medical expertise about the essentiality of the device. Conformity to the essential safety and performance requirements. Approval or certification in
other industrial states worldwide. Experiences in human use / animal studies. Current, completed or planed clinical trials.
• An explanation of the potential risks for the patient, a consent of the patient, stating the patient agrees to the use of the non-certified medical device, a statement declaring there is no suitable alternative in Europe.
• The medical reason for the need of the device; amount; alternatives; previous CE status, etc.
• Same standard request template as national derogation procedure.
• Additional explanations are requested case by case.
• If an equivalent CE marked medical device is available on the Union market, an explanation on what differentiates the non-CE marked device from equivalent CE marked Devices, the clinical benefit- risk assessment for the patient using the device.
• name of the required medical device, including any accessories+justification of the extent to which the use of the medical device is warranted in the interest of health protection+evidence of the lack of availability of equivalent medical devices for which the conformity assessment procedures have already been carried out+confirmation of compliance with the essential safety and performance requirements applicable to the medical device, taking into account its intended purpose. Where certain essential safety and performance requirements are not fully complied with, a detailed benefit/risk assessment demonstrating that the use of the medical device is appropriate despite this partial non-compliance and that, with regard to the requirements not fully complied with, all necessary measures have been taken to protect patients, users and, where applicable, third parties.
• It is the same as for the other derogations.
• Are there similar CE-marked devices? / If so, why can’t they be used? / Are there alternative treatments (medical devices, medicines or other treatments)? / If so, why can’t they be used? / Has the device been approved by the FDA? / If so, what is the intended purpose of this device ? / Is the device undergoing clinical investigation - performance study? / If so (ongoing or finished), What is the study title ? / If so (ongoing or finished), What is the complete scope ? / If the clinical investigation / performance study is still ongoing, is it located in Belgium ? / If so, why can’t the patient / patient sample be included in the study ? / A special form has to be filled-in by a physicien (sex, age, medical condition, medical reason for the application / consequence if the device is not used, date of the surgical intervention, is there similar devices or treatment, why they can't be used)
What supporting documents are you asking for...
Other supporting documents • A request under article 59 MDR must contain all the following elements, identification of the hospital(s) and physician(s)
responsible for the use of the device, duly justified grounds for the delivery of the non-CE marked device, a medical justification (by the hospital) for the use of the medical device in question, including the nonexistence of an alternative CE marked device, and/or the non-suitability of alternative techniques/procedures, authorization from the Executive Board of the hospital(s) where the procedure will take place, the opinion of the Health Ethical Committee of the aforementioned hospital(s)
• Patient consent
• A copy of the chamber of commerce, an explanation of the potential risks for the patient, a consent of the patient, stating the patient agrees to the use of the non-certified medical Device, a statement declaring there is no suitable alternative in Europe, a doctors statement
• Risk analysis, countries where placed on market, statement from healthcare professional
• Supporting documentation may be tailored to the requirements of the specific application
• The assessment is made on a case-by-case basis, the most important thing is that the applicant provides us with all the safety guarantees available
• Depending on the case, healthcare professionals can help to justify the public health interest involved
• Application is relevant to a non CE marked device, documents verifying the performance of the non-CE marked device, if the concerned device includes materials for manufacture which are derived from animal origin (excluding devices which contain material of animal origin which are externally applied and are not placed in contact of broken skin)
Does the healthcare institution/professional user also justify the derogation?
Do you want to know that the applicant has informed the patient of the non-compliance?
If you answered yes to the previous question, do you require as proof either of the following:
If it is an urgent request situation for compassionate/single patient use do you apply...
Urgent request situation
• The case gets a higher prioritization, but the standard procedure will apply as well
• In very rare case we grant a compassionate use after the surgery, in this case, the applicant must demonstrate that it was not possible to fill-in an application before the surgery due to the emergency
• Urgent request situations require an individual risk benefit assessment, medical opinion on the essentiality of the device
• The special procedure, less documents, not full tehnical documentation
• The health care professional's statements is most requested
• Standard procedure is completed urgently in these cases
• CA initiates on its own motion necessary measures to clarify the case sufficiently, an explanation why the use of a medical device is vital, details on vigilance or recall issues to date, substantiation to support the risk/benefit-analysis
• Informing the patient, in cases of vital emergency, authorization is granted on the basis of a template mail and restricted elements
• Request for information on the absence of alternatives available in view of the life-threatening emergency to be identified (ex : artifial heart)
Have you established a minimum number of days between the submission of your application and the decision on the application?
Do you also ask the applicant to confirm or take responsibility that it is a non-compliant medical device?
Approximately how many compassionate/single patient use derogation requests do you receive/approve/reject per year?
Country Receive Approve Reject
1. - - -
2. 8 8 0
3. 100-150 App 90% App 5%
4. 1 0 0
5. 3 2-3 1
6. 2-3 2-3 0
7. 25 24 1
8. 2 2 0
9. - - -
10. 15 10 0
Approximately how many compassionate/single patient use derogation requests do you receive/approve/reject per year?
Country Receive Approve Reject
11. 2 2 0
12. Compassionate use/ single patient use is not defined in EU legislation or in national law.
13. 36 26 2
14. - - -
15. 0 0 0
16. 6 5 1
17. 0 - -
18. 3-5 3-5 0
19. 63 62 1
If rejected, what is the most common reason?
Have you had cases where the applicant has either withdrawn the application or failed to complete the process?
Reason for withdrawal or failure to complete process
• Not all required elements were submitted by the applicant
• Different
• We do not know
• Compassionate Use not applicable
• Wrong department (did not turn out to be compassionate Use)
• Adequate documentation was not provided to enable adequate assessment
• Death or use of another emergency technique, when the request comes from the healthcare
professional but the manufacturer does not wish to follow this procedure
Fees
Country Fee amount
1. Currently, no
2. 250 - 10.300 EUR
3. 500 Euro
4. 193 Euro per hour
5. 62000SEK
6. The fee varies between 1930€ - 4070 €
Useful information/experience
• Usually authorisations under article 59 MDR are under the premise "one device (unit), one hospital/patient". Exceptionally, authorisations
are granted to several units of one device for one hospital. We do not grant national-wide authorisations, as we consider these to be
exceptional situations and also because, as we have already referred in one of the previous answers, the manufacturer must submitted
elements which are produced by each hospital where the non-CE marked device will be used
• Art. 59 is essential because it makes it possible to treat patients with not certificated devices based on individual cases and to act against
device supply shortages. Applicants and users reveal a high grade of responsibility using the possibilities of Art. 59
• Application forms really help with setting expectations as well as rising the bar for an application just enough so that only serious
applications come in
• Important national competence and there is a large expectation in our health system for such a service, there are considerable resource
expertise and governance considerations for CA's processing derogations
• Processing of the requests is found to be very time consuming and laborious
• It is almost impossible to identify equivalent medical devices which are CE-marked (we do not expect EUDAMED to solve this problem
utterly)
• Ggood experiences, some difficult questions in the context of manufacturer not willing to receive CE-marking for their products (MDR).
Future plan -What do you think should be changed in the derogation
procedure?
• Filling the CIRCABC derogation table is sometimes problematic, because the table is sometimes checked out
an other states are not able to deliver the data any more
• An established procedure and a specific form for this purpose with both derogations
• In the process of aligning our internal procedure, on EU level, maybe there should be a guidance to ensure
uniformity of handling and assessment of the applications and that outlines "interest of public health or patient
safety or health"
• Optimal application of the procedure would benefit from greater visibility with regards expectations and
documents expected between MS’s, in addition, there is currently no clear visibility or established procedure for
extending a national derogation to Union wide level under A59(3), would like to review the use of derogations
with other MS’s and explore the most appropriate use of A59(3), for example in a coordination call
• Currently it is free to apply for a derogation, handling derogation invest a lot of time and resources, it should be
the manufacturer/ applicant who pays for the deviation from conformity, not the general tax payer
Future plan -What do you think should be changed in the derogation
procedure?
• It should be harmonised for all CAs
• The application form should be updated to save time during the examination phase
• Process development is ongoing
• Not sure that 100% of the derogation are uploaded on Circa, would be useful to also upload refusal and the
reasoning behind, it could be done via a modification of the art. 59.2. due to the “notified body bottleneck” there
are more and more derogation request due to a too long certification procedure are shortages on the market, it
was not the initial goal of the art. 59 and in some cases it could be seen as a shortcut to the certification
process, eudamed was foreseen as a central tool of the legislation, it should have been useful for derogation,
for example via an easy access to the certificate and a list of the similar devices that are on the market, need to
have Eudamed as soon as possible, it seems that the criteria’s used to grant a derogation vary among EU
Competent authorities, the duration of the derogation varies in each country
• Consider to introduce a common standard EU/EEA-template for the derogation process
Thank you for replying Your input was very helpful
Chief Specialist
Department of Medical Devices
Estonian Health Board
Maarika Ojala